From: Conduct of a personal radiofrequency electromagnetic field measurement study: proposed study protocol

Figure 2

Functional test of exposimeters. Example of functional tests of the GSM900 downlink band of the six devices used in a Swiss study (QUALIFEX) conducted by the Federal Office of Metrology in Wabern, Switzerland. All relative changes refer to the V/m units. The tests revealed a problem with device number 4 in December 2008. All other changes were within the measurement uncertainty of ±2 dB.